One Probe, Four Readings: Milwaukee MI805 pH/EC/TDS/Temp Meter
The Milwaukee MI805 reads pH, conductivity, TDS and temperature from one MA851D/1 amplified probe. That probe puts all four sensors in a single rugged handle on a 1 m cable with a DIN connector. pH runs from 0.00 to 14.00 at 0.01 resolution, with accuracy of ±0.01 pH at 25 °C. Conductivity covers 0 to 3999 µS/cm and TDS covers 0 to 1999 ppm, both at ±2% of full scale. Temperature reads 0.0 to 60.0 °C or 32.0 to 140.0 °F, and the scale is user selectable. Automatic temperature compensation works across that span. The compensation coefficient adjusts from 0.0 to 2.4% per °C. The EC to TDS conversion factor adjusts from 0.45 to 1.00 to match the ppm scale a feed chart uses. A stability indicator flags the moment a reading settles. Auto-hold freezes the value, so it can be logged after the probe comes out of the tank. The large display shows pH with temperature, or EC and TDS with temperature, side by side. One 9V alkaline battery runs about 300 hours, and the meter shuts off after 8 minutes idle.
TDS pH Meter Combo Calibration and Reservoir Checks
Calibrate pH first, at the start of a feed day. Pick the buffer set in setup, either 7.01 with 4.01 and 10.01, or the NIST set at 6.86 and 9.18. Remove the cap, set the probe in the first buffer and hold the power key until CAL shows. The meter recognizes the buffer on its own, then asks for the second one. Rinse the probe before EC calibration, then dip it in 1413 µS/cm solution and hold the same key. The meter calibrates automatically and confirms with OK on the screen. For a reservoir check, stir the tank, dip the probe tip in the sample and wait for the stability indicator. Read pH, then press RANGE to read EC or ppm from the same dip. Check runoff the same way to see whether salts are building in the media. Keep a few drops of storage solution in the cap between uses. A dried tip recovers after soaking the bottom 2.5 cm in rinse solution for a few minutes.
Key Features
- One Probe, Four Parameters: The MA851D/1 amplified probe reads pH, EC, TDS and temperature in one dip.
- Automatic Buffer Recognition: One or two-point pH calibration with standard or NIST buffer sets.
- Automatic EC Calibration: One-point calibration in 1413 µS/cm solution.
- Adjustable TDS Factor: Set from 0.45 to 1.00 to match the ppm scale in use.
- Adjustable ATC Coefficient: Beta set from 0.0 to 2.4% per °C for repeatable conductivity readings.
- Stability Indicator and Auto-Hold: Shows when a reading settles and freezes it for logging.
- Long Battery Life: About 300 hours on one 9V alkaline battery, with auto-off after 8 minutes.
- Complete Kit: Calibration and cleaning sachets, a 9V battery and a rugged carrying case.

Technical Specifications
| Specification | Value |
|---|---|
| Brand | Milwaukee Instruments |
| Model | MI805 |
| Parameters | pH, EC, TDS, temperature |
| pH Range | 0.00 to 14.00 pH |
| EC Range | 0 to 3999 µS/cm |
| TDS Range | 0 to 1999 ppm |
| Temperature Range | 0.0 to 60.0 °C / 32.0 to 140.0 °F |
| Resolution | 0.01 pH; 1 µS/cm; 1 ppm; 0.1 °C / 0.1 °F |
| Accuracy (@25 °C) | ±0.01 pH; ±2% full scale EC/TDS; ±0.5 °C / ±1 °F |
| Temperature Compensation | Automatic, 0 to 60 °C, β adjustable 0.0 to 2.4%/°C |
| pH Calibration | Automatic, 1 or 2-point, automatic buffer recognition |
| EC Calibration | Automatic, 1-point |
| EC/TDS Conversion Factor | Adjustable 0.45 to 1.00 |
| Probe | MA851D/1 amplified pH/EC/TDS/temperature probe, DIN connector, 1 m cable |
| Environment | 0 to 50 °C / 32 to 122 °F, max. 100% RH |
| Battery | 1 x 9V alkaline, approx. 300 hours |
| Auto-Off | After 8 minutes of non-use |
| Dimensions | 200 x 85 x 50 mm |
| Weight | 260 g with battery |
| Included | MA851D/1 probe, 2 x 20 mL pH 4.01 and pH 7.01 sachets, 2 x 20 mL 1413 µS/cm sachets, 2 x 20 mL cleaning sachets, carrying case, 9V battery, instructions |